Start Up

The Color Measurement Menu is displayed when diagnostics are
completed. Insert a cuvette/sample cell with distilled water to
calibrate.The calibration starts automatically when the LCM III
recognizes a 11 mm round vial in the cell compartment #1. For the
use of square cuvettes/sample cells in the cell compartment #2,
press Start to start the calibration.

Insert the sample cuvette/cell. For the use of 11 mm round vials the
measurement starts automatically. For the use of square
cuvettes/sample cells in the cell compartment #2, press Measure to
start the measurement.

See section

5.1.3 Main Menu on page 16

for a detailed description.

4.4 Characteristics in continuous operation

After a continuous operating time of more than 24 hour without
switching off the instrument we recommend to perform a new
calibration with a distilled water cuvette/sample cell.

After a continuous operating time of more than 7 days without
switching off the instrument we recommend to switch off and on the
instrument to perform a new system check with filter adjustment
and lambda-calibration.