background image

40

Advanced Operations

6.

Remove any cuvettes/sample cells from the cell compartment
and press Start.

7.

Insert the different filter in the given order one after the other.
Press Next after inserting a filter.

After the last measurement the results are displayed.

8.

Press PC & Printer icon to send the data to a USB memory
stick, PC or to a printer.

The files will be stored automatically as CSV file (Comma
Separated Value). The file name will be formatted as
“Verification.csv“.

6.1.4 Output checks

If a printer is connected a test printing of the current screen will be
printed.

6.1.5 Lamp history

The Lamp History menu provides the amount of time that the lamp
has been on (Hours).