ALTANA Spectro-Guide Sphere Gloss Manual User Manual

Page 22

Advertising
background image

21

Calibrate

chk green

To determine whether recalibration of the instrument
is necessary, perform a test measurement with the
green standard. From the “Option” menu, use the

button to place the cursor on the “Calibr.” option.

When you press the Operate button, the submenu
opens. The option “chk green” is preselected.

Take the green test standard and open the lid.
Position the measuring instrument so that the
display points in the direction of the open lid and the
front edge of the instrument is parallel to the edge of
the lid. Press the Operate button. You will then be
instructed to measure the green test standard.

A message will then appear, indicating whether the
calibration is still valid. If the message

Please calibrate

appears, you need to perform the calibration with
the black and white standards.

std

File

Display

Color sys
Index

Illum/Obs

Setup

Calibrate

Configure

f

f

f

f

f

Option

Calibr.

Chk Green

f

Gloss

f

File

Display

Measure Green Std

Press Operate Button

Option

File

Display

Calibration OK

Option

Clean the standards with a moist cloth. Never use
window cleaner or any abrasive cleaning agents. For
stubborn stains, saturate the cloth with ethanol.