ALTANA Spectro-Guide Sphere Gloss Manual User Manual

Page 23

Advertising
background image

22

File

Display

Option

Measure Gloss Std

Press Operate Button

Std.No. : 976802

Calibrate

Use the

buttons to move the cursor to the “Option” menu.
From this menu, use the

button to place the cursor on the “Calibr.” option.
Press the Operate button. When the submenu
opens, use the

button to move the cursor to the “Calibrate” option.
Now press the Operate button, and calibration will
begin. The instrument will first call for the black
standard. Place the instrument on the black
standard and press the Operate button.

A beep will indicate that the measurement has been
completed.

Now place the instrument on the white standard and
press the Operate button.

For gloss calibration, place the instrument on the
high-gloss standard and press the Operate button.

If calibration was performed correctly, the message

Calibration OK

will appear on the display. The instrument returns to
the previous measurement mode.

Use the test with the green standard again to
determine whether the calibration was performed
correctly.

std

std

std

File

Display

Color sys
Index

Illum/Obs

chk green

Setup
Configure

f

f

f

f

f

Option

Calibr.

Calibrate

f

Gloss

f

File

Display

Option

Measure White

Press Operate Button

Std No 638444

File

Display

Color sys
Index

Illum/Obs

Setup
Configure

f

f

f

f

f

Option

Calibr.

f

Gloss

f

File

Display

Option

Measure Black Trap

Press Operate Button