10.5 Calibration

Chk Green

From the “Options” menu, use the

button to move the cursor to the “Calibr.” option.

The submenu opens when you press the Operate
button. The “chk green” option is preselected.

Take the green test standard and open the lid.
Position the instrument so that the display points in
the direction of the open lid and the front edge of the
instrument is parallel to the edge of the lid. Press
the Operate button.

At this point, you will be prompted to measure the
green test standard.

You will then see a message indicating whether the
calibration is still valid. If the message

Please calibrate!

appears, the calibration must be performed with the
black and white standards (see Chapter 6 –
Calibrate).