Options

Opacity

Opacity can only be performed in Difference mode.
To activate this function, move the cursor to the
“Opacity” option as described for the “Options/Index”
menu. The instrument then returns to the measuring
mode. All values on the display are set to zero. To
determine the opacity of a sample, use the

button to measure the sample over the black
background. Then use the Operate button to
measure the sample over the white background.

After measuring with the “Std” button, the opacity is
zero. The L*, a*, and b* values for the sample over
the black background are displayed on the right
side.

After the measurement has been performed with the
Operate button, the opacity and the delta values
between the measurements over the black and
white backgrounds are displayed.

If you would like to store the reading with the “File/
Save/Sample” function, only the reading against the
white background is stored.

std

92.45%

File

Display

Option

Opacity

L*

91.98

a*

0.41

b*

-1.38

E*

10.54