47

10.6.3 Metamerism index limit

From the “Options” menu, use the

button to move the cursor to the “Setup” option.

The submenu opens when you press the Operate
button.

Move the cursor to the “MI Limit” option and press
the Operate button. The input menu for the MI limit
appears. Use the

buttons to change the underlined digit. Use the

button to move the underline to the next digit.

End the input operation by pressing the Operate
button. The instrument returns to the measuring
mode.

To end the input without storing, use the

buttons to exit the input field.

The metamerism index is now calculated for each
difference measurement. If it is greater than the
selected limit, a reverse “M” appears on the display.
To switch off the calculation, the limit must be reset
to 0.00.