8.2.1 Calibration of temperature value

During temperature calibration only a check is carried out, i.e. no values are changed.

 Invoke service function „Tmp test“, see chap. 8.2

 Confirm by pressing the PRINT button. The current setting will be displayed. Use

the navigation buttons

  to select the temperature during the check, available

35- 160°C, e.g. 120°C.

 Confirm using the PRINT button, the appliance heats up to the set temperature.

 After approx. 15 minutes compare the test temperature with the displayed

temperature of the DLB-A01N. If the two values do not match, we recommend a
temperature adjustment, see chap. 8.2.2

 Use the ON/OFF button to finish the calibration.

 Use the ON/OFF button to return into weighing mode.