11 rs 232c interface, 1 technical data, 2 pin allocation of the output plug – KERN & SOHN DLB 160-3A User Manual

Page 46

Advertising
background image

46

DLB_A-BA-e-1511

11 RS 232C interface


The moisture analyser is typically equipped with a RS 232C interface.
The following conditions must be met to provide successful communication between the
moisture analyser and the printer.

• Use a suitable cable to connect the moisture analyser to the interface of the

printer. Faultless operation requires an adequate KERN interface cable.

• Communication parameters (baud rate, bits and parity) of moisture analyser and

printer must match.

• Data may only be transferred in moisture analysis mode.

The measurement data may be edited according to the setting in the menu item „Serial“
(see chap. 11.3) or automatically or by pressing the PRINT button via the interface.

11.1 Technical data

 8-bit ASCII Code
 8 data bits, 1 stop bit, no parity bit
 Baudrate selectable from 1200 - 9600 Baud , see chap. 9.1.1
 For operation with interface faultless operation is only ensured with the correct

KERN – interface cable (max. 2m)

11.2 Pin allocation of the output plug




Pin 2:

Tx Signal

Pin 3:

Rx Signal

Pin 5:

GND