KERN & SOHN DLB 160-3A User Manual

Page 24

Advertising
background image

24

DLB_A-BA-e-1511

Menu overview

Menu item

Display

Selection

Description

Baudrate
(siehe Kap. 9.1)

bAud rt

br 1200

br 2400

br 4800

br 9600

Automatic zero point
correction (see chap. 9.2)

Auto 0

Au0 OFF

Auto Zero switched off

Au0 1

Auto Zero range ± ½ digit

Au0 2

Auto Zero range ± 3 digits

Au0 3

Auto Zero range ± 7 digits

Au0 3E

Auto Zero range ± 7 digits in the
whole weighing range

Filter
(see chap. 9.3)

FiltEr

Filt 1

Setting for dispensing

Filt 2

Sensitive and fast, very quiet
set-up location.

Filt 3

Robust but slow, busy set-up
location.

Stability display
(see chap. 9.4)

StAbil

Stab 1

Standstill control fast / very quiet
set-up location

Stab 2

Standstill control fast + exact/
quiet set-up location

Stab 3

Standstill control exact /very set-
up location.

Background illumination
of the display (see chap.
9.5)

Blt

on

Background illumination on

off

Background illumination off

Auto

Backlighting automatically
switched off 3 seconds after
achieving stable weighing value.
Changes in weight or pressing
of keys will automatically result
in backlight switching on again.

End