15 instant help – KERN & SOHN DLB 160-3A User Manual

Page 58

Advertising
background image

58

DLB_A-BA-e-1511

15 Instant help

Fault

Possible cause

Display is not lit up.

• The display unit is not switched on.

• The mains supply connection has been

interrupted (mains cable not plugged in/faulty).

• Power supply interrupted.

Measurement is taking too
long

• Incorrect setting shutoff criterion

Measurement is
not reproducible

• Sample is not homogenous

• Drying time is too short

• Drying temperature too high (e.g. oxidation

sample material, boiling point of sample
exceeded)

• Temperature sensor soiled or defective

The displayed weight is
permanently changing

• Draught/air movement

• Table/floor vibrations

• Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)