1 how to implement drying – KERN & SOHN DLB 160-3A User Manual

Page 32

Advertising
background image

32

DLB_A-BA-e-1511

10.1 How to implement drying

Place removal aid with empty sample dish on the sample dish retainer. Make sure that
the sample dish is resting flat on the sample dish retainer. Use the sample retainer at all
times as it allows safe working and prevents burns.

Prior to start the moisture analysis select a suitable drying program.

 In weighing mode press the MENU button, „Therm“ is displayed.

 Acknowledge using PRINT button, „Measure“ is displayed.

 Acknowledge using PRINT button, the current setting is displayed.

 Use the navigation buttons   to select the desired drying program.

s. chap. 10.1.1

s. chap. 10.1.2

s. chap. 10.1.3