13 error messages – KERN & SOHN DLB 160-3A User Manual

Page 56

Advertising
background image

56

DLB_A-BA-e-1511

Material

Weight

Sample

(g)

Drying

temperature

( °C)

Drying

period

(approx.)

(min)

Moisture

%

(approx.)

Solid

body

%

(approx.)

Tobacco

1,5

100

16

10,18

Tea, black

2

105

4

7,67

Pasta

1,5

120

8

10,64

Textile fibre

0,8-1,2

85

3,6

14,03

Theophylline

1,5

130

1,9

7,33

Thermoplastic PUR –

granulate

15-18

80

18

0,08

Walnut

2,8

100

5,6

3,5

Washing powder

2

160

12

7,32

Wheat spring water

2-3

90

10

6

Sausage casing

0,2

150

3,5

78,56

Toothpaste

2

100

7,7

34,28

Pulp

2,5

130

4,5

7,32

Cement

8-12

138

4-5

0,8

Sugar

4-5

138

10

11,9

Sugar beets

2

130

13,4

30,94


13 Error messages

ERR01

Weight value instable or zeroing not possible.
Check the environmental conditions.

ERR02

Adjustment error, e.g. instable environmental conditions

ERR03

Adjustment error e.g. incorrect adjustment weight}

ERR05

Data transfer not possible, as weighing value is instable. Check
the environmental conditions.

ERR07

Faulty data reading

ERR10

Instable display when drying starts, check environmental
conditions

ERR11

Sample weight too small

UNLOAD”:

Sample or sample dish wrongly positioned.

Err thb

Heating top does not work, check current supply

Weighing range exceeded, placed load exceeds the capacity of
the appliance. Unload appliance.

Weighing range not reached, e.g. Dish holder / removal help is
missing.

Advertising