ALT_B-PLT_A-PLT_F-BA-e-1333

7.3 Adjustment with external adjustment weight

 Condition:

Menu setting „External adjustm.“, see chap. 7

 At the models with internal adjustment weight the adjustment with

external weight is not recommended.

 Information about test weights you will find in the internet under

http://www.kern-sohn.com

 Ensure that there are no objects on the weighing pan. In weighing mode press

the CAL button.

 Wait until the weighed value for the required adjustment weight appears

flashing.

During the flashing display put the required adjustment weight carefully in the

center of the weighing pan. The flashing display disappears.

 After successful adjustment the balance automatically returns to weighing

mode.

 Take away adjustment weight