7.4 Overwrite internal adjustment weight

 Overwriting is restricted to specialist staff possessing well acquainted with

the workings of weighing scales.

Information about test weights you will find in the internet under
http://www.kern-sohn.com

Call up menu item „Techn. adjustment“, see chap. 7.

 To confirm press the PRINT button and keep it pressed until the acoustic signal

gets mute.

 Press the MENU button and keep it pressed until the acoustic signal gets mute.

The balance returns automatically into weighing mode.

 Ensure that there are no objects on the weighing pan. Press the CAL key

 Wait until the weighing value of the adjustment weight appears flashing.

During the flashing display put the required adjustment weight carefully in the

center of the weighing pan. The flashing display disappears.

 Wait for stability display, remove adjustment weight.

 Keep the PRINT button pressed until the acoustic signal gets mute. The internal

adjustment weight is overwritten.

 After successful overwriting, the balance automatically returns to weighing

mode.