P. 11.7), 6 stability – KERN & SOHN ALT 250-4B User Manual

Page 45

Advertising
background image

ALT_B-PLT_A-PLT_F-BA-e-1333

45

11.6 Stability

 In weighing mode press the MENU button and keep it pressed until the acoustic

signal gets mute. The setup menu is displayed.

 Use the navigation buttons 

to select the menu item „Stability“.

 Acknowledge using PRINT button, the current setting is displayed.

 Use the navigation buttons  to select the desired setting.

Display

Description

Stability 1

Standstill control fast - very quiet set-up location

Stability 2

Standstill control fast + exact

– quiet set-up location

Stability 3

Standstill control exact / set-up location.

 Take over selection using the PRINT button.

Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.

 Press the MENU button and keep it pressed until the acoustic signal gets mute.

The balance returns automatically into weighing mode.