8.1 Automatic adjustment with internal weight

The automatic adjustment with internal weight

is started

automatically, if the balance

after the weighing balance was disconnected from the mains

after pressing ON/OFF in stand-by mode

after a temperature change of 1.5 °C

with non loaded weighing pan / zero display

after a time interval of 20 min with non loaded weighing pan / zero

display

The automatic adjustment function is always enabled. At any time, however, you can
carry out a manual adjustment (press CAL-button) using the internal adjustment
weight, see chap. 7.1.

Or selecting the adjustment mode in the menu:

 In weighing mode press the MENU button and keep it pressed until the acoustic

signal gets mute. The setup menu is displayed.

 Use the navigation buttons 

to select the menu item „Adjustment mode“.

 Confirm by the PRINT button

, „Autom. Just.“ will be displayed.

 Take over selection using the PRINT button.

Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.

 Press the MENU button and keep it pressed until the acoustic signal gets mute.

The balance returns automatically into weighing mode.