P. 11.13), 12 adjustment mode – KERN & SOHN ALT 250-4B User Manual

Page 51

Advertising
background image

ALT_B-PLT_A-PLT_F-BA-e-1333

51

11.12 Adjustment mode

 In weighing mode press the MENU button and keep it pressed until the acoustic

signal gets mute. The setup menu is displayed.

 Use the navigation buttons 

to select the menu item „Calibration mode“.

 Acknowledge using PRINT button, the current setting is displayed.

 Use the navigation buttons  to select the desired setting.

! In models PLT-M only automatic adjustment is available.

Display

Description

Procedure see chap.

Internal

adjustm.

Adjustment with internal
weight after pressing the
CAL button

+

Chap. 7.1

External

adjustm.

Adjustment with external
adjustment weight

+

Chap. 7.3

Technical

adjustm.

Overwrite internal
adjustment weight

+

Chap. 7.4

Autom. adjustm.

Automatic adjustment with
internal adjustment weight

+

Chap. 7.2

+

or 8.1 (PLT-M)

 Take over selection using the PRINT button.

Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.

 Press the MENU button and keep it pressed until the acoustic signal gets mute.

The balance returns automatically into weighing mode.

Advertising