P. 11.6), P. 11.5) – KERN & SOHN ALT 250-4B User Manual

Page 44

Advertising
background image

44

ALT_B-PLT_A-PLT_F-BA-e-1333

11.5 Filter

This menu item allows the balance to be set according to specific ambient conditions
and measuring purposes.

 In weighing mode press the MENU button and keep it pressed until the acoustic

signal gets mute. The setup menu is displayed.

 Use the navigation buttons 

to select the menu item „Filter“.

 Acknowledge using PRINT button, the current setting is displayed.

 Use the navigation buttons  to select the desired setting.

Display

Description

Filter 1

Setting for dispensing

Filter 2

The balance reacts quickly and in a sensitive manner,
quiet set-up location.

Filter 3

The balance reacts slowly and in a robust manner,
busy set-up location

 Take over selection using the PRINT button.

Weighing balance returns to menu. Either make more settings in the menu or go
back to weighing mode as follows.

 Press the MENU button and keep it pressed until the acoustic signal gets mute.

The balance returns automatically into weighing mode.

Advertising