1 crucial features of performance, 4 minimum distances – KERN & SOHN MPE 250K100HM User Manual

Page 21

Advertising
background image

8.3.1 Crucial features of performance

Note:

The floor scales MPC/MPE does not have any crucial features of
performance as per IEC 60601-1. The system may be subject to
interference by other devices even if these devices conform to current
emission requirements as per CISPR.

8.4 Minimum distances

Recommended safety distances between portable and mobile HF
telecommunication devices and the medical device

The floor scales MPC/MPE is designed for use in an electromagnetic environment in which HF
disturbance variables are controlled. The customer or user of the medical electrical device can help
avoiding electromagnetic disturbances by keeping the minimum distance between portable and mobile
HF telecommunication devices (transmitters) and the floor scales MPC/MPE – depending on the
output performance of the communication device, as stated below.

Rated capacity of

transmitter

W

The safety distance depends on the transmission frequency

m

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.20

1.20

2.30

10

3.80

3.80

7.30

100

12.00

12.00

23.00

For transmitters with a maximum rated capacity not stated in the table above you can calculate the
recommended safety distance in metres (m) yourself by using the equation belonging to each column,
whereby P equals the maximum rated capacity of the transmitter in Watt (W) as per details provided
by the transmitter manufacturer.

NOTE 1 Higher frequency range applies to 80 MHz and 800 MHz.
NOTE 2 These guidelines may not be applicable in all cases.

The spread of electromagnetic variables is influenced by absorption and reflections in

buildings, objects and humans.



MPC_M / MPE_HM / MPE_PM-BA-e-1414

21