12 the menu, 1 navigation in the menu – KERN & SOHN MPE 250K100HM User Manual

Page 38

Advertising
background image

12 The menu

Access to service menu „tCH“is locked in verified weighing scales.

To disable the access lock, destroy the seal and actuate the
adjustment switch. For position of adjustment switch, see chap. 16.

Attention:
After destruction of the seal the weighing system must be re-verified
by an authorised agency and a new verification wire/seal mark fitted
before it can be reused for applications subject to verification.


12.1 Navigation in the menu

Call up menu

In weighing mode, press

and the first function [F1

oFF] will be displayed.

Select function

With help of

, the individual functions can be

selected one after the other.

Change settings

Confirm selected function by pressing

(MPE) or

(MPC). The current setting will be displayed.

To select the desired setting, press

and confirm

your selection by

(MPE) or

(MPC); the

weighing scale will return to the menu.

Exit menu/

Return to weighing
mode

Press

and the scales will return to weighing mode.



38