13 Data output RS 232


You can print weighing data automatically via the RS 232 interface or manually by

pressing mit

(MPE) bzw.

(MPC) via the interface according to the setting

in the menu.

This data exchange is asynchronous using ASCII - Code.

The following conditions must be met to provide successful communication between
the weighing balance and the printer.

• Use a suitable cable to connect the weighing balance to the interface of the

printer. Faultless operation requires an adequate KERN interface cable.

• Communication parameters (baud rate, bits and parity) of balance and printer

must match. For detailed description of interface parameters see chap.

12.1)

In a medical context only auxiliary equipment in compliance with Directive
EN 60601-1 may be connected to the interface.

13.1 Pin allocation of balance output bushing:



Pin 2: TXD - Output
Pin 3: RXD - Input
Pin 5: GND - Signal ground

13.2 Technical data

Connection

9 pin d-subminiature bushing

Pin 2 output

Pin 3 input

Pin 5 signal earth

Baud rate

Optional 600/1200/2400/4800/9600

Parity

8 bits,



MPC_M / MPE_HM / MPE_PM-BA-e-1414