KERN & SOHN MPE 250K100HM User Manual

Page 54

Advertising
background image

KERN MPE

Press

,

and

one after the other and [P1 SPd]

will appear

KERN MPC

Press

,

and

one after the other and [P1 SPd] will

appear

Press

, „P2 CAL“ will be displayed

Operate adjustment switch; for position see chap.16

Press

(MPE) or

(MPC) and [dESC] will appear

Press

repeatedly

until „CAL“ will be displayed.

To confirm, press

(MPE) or

(MPC) and [UnloAd]

will appear

Ensure that there are no objects on the weighing pan.

Wait until the stability display „STABLE“ is displayed and then

confirm by pressing

(MPE) or

(MPC).

(example)

The size of the currently set adjustment weight will be
displayed.

To change, select the digit to be altered by

, and the

numerical value by

.

Confirm by pressing

(MPE) or

(MPC) and [LoAd]

will be displayed.