7.3.9 Titrant Concentration Parameter

When mode is generated, the parameter is usually set to BLANK and user should input this
parameter when calculating concentration value (see 8.1.3 ‘Sample Concentration).


7.4 Blank Titration Mode

This mode is suitable for the titration that demands less titrant (below 1ml). In this mode, every
time the meter adds 0.02 ml of volume (users can amend this parameter), users can also set pre-
added volume parameter so as to make titration speed higher. Thus the meter will automatically
search for titration end point to generate special titration mode. When the volume is small in
titration, user can select blank titration. In starting status, press ‘Titration’ button, select ‘Blank
Titration’ to enter into blank titration mode, displayed as follows:


Start Titration

Stirring

Resume set

Pre-adding Volume: 0.5mL

Each Adding: 0.02mL

Ending Volume: 5mL

Balance Time Switch: ON
Value: 10s

Balance Potential Switch ON
Value: 0.2mV


MV Titration pH Titration


First end point 100

MV/pH Titration

Jump Value Set

Blank Titration/Parameter setting









Diag.35 Parameter Setting in Blank Titration Mode

Blank titration mode includes: mV/pH Titration Parameter, Balance Potential Switch and Value

Parameter, Balance Time Switch and Value Parameter, End Point Jump Value Parameter,

Pre-adding Volume Parameter, Each Adding Volume Parameter, End Volume Parameter.

7.4.1mV/pH Titration Parameter

The same as mV/pH Titration Parameter in Pre-Titration Mode.


7.4.2 Balance Potential Switch and Value Parameter

The same as Balance Potential Switch and Value Parameter in Pre-Titration Mode.

7.4.3 Balance Time Switch and Value Parameter

The same as Balance Time Switch and Value Parameter in Pre-Titration Mode.