Vi. the complete sets of the meter, Vii.appendix – LABEC M-TIT-5 User Manual

Page 48

Advertising
background image

VI. THE COMPLETE SETS OF THE METER

1. Model

T I T-5 Automatic Potentiometer 1 set

2. Model

TIT-5 Automatic Potentiometer volume titration and measuring unit 1

3. Model T-818-B-6 Temperature Sensor

1

4. Model E-201-C-9 pH Combination Electrode

1

5. Accessories,

see

Packing

List

1

VII.APPENDIX

1. pH standard buffer solution A (pH 4.00, 25℃)

Weigh 10.12g of KHC

8

H

4

O

4

that has been dried under 110℃~130℃ for 2~3 hours, dissolve it in

1L of de-ion water.

2. pH standard buffer solution B (pH 6.86, 25℃)

Weigh 3.388g of KH

2

PO

4

and 3.533g of Na

2

HPO

4

respectively, dissolve it in 1L of de-ion water.

3. pH standard buffer solution C (pH 9.18, 25℃)

Weigh 3.80g of Na

2

B

4

O

7

, H

2

O which was put in the desiccator with saturated NaBr for two days,

dissolve it in 1L of de-ion water.


Table 1

Symptom Trouble

cause Settlement

No display


a. power is off
b. fuse is damaged
c. bad connection between

control unit and volume
titration unit

a. check the power
b. change the fuse of same

type

c. check the connection

between control unit and
volume titration unit

Connecting failed when switch
on the meter

a. connection between each

unit is incorrect.

b. reduction of the meter is

sometimes no good

a. check the connecting line

according to 1.2 in chapter
IV Operation of the Meter

b. switch off and switch on

again after 30 seconds

mV/pH measurement is not
right

a. something wrong with

electrode performance

b. something wrong with

another electrode interface
short circuit

a. change an electrode
b. change Q9 short circuit

plug

Printer doesn't work or work
incorrectly

a. power of printer is not

connected

b. print line is not connected
c. printer is set by mistake
d. printer is selected by

mistake

a. connect printer power
b. connect print line
c. set correct printer type

according to 10.3

d. change the printer

Pre-titration can not find the
end point

a. end point jump is too small
b. titrant or sample are not

right

c. end point value is too small
d. electrode is selected by

mistake

a. set jump to "small"

according to 7.1.5

b. change titrant or correct the

sample

c. use "blank titration" mode
d. select electrode correctly

47