LABEC M-TIT-5 User Manual

Page 43

Advertising
background image

10.3 Set Printer

RS-232 interface is taken as output of the meter. Therefore, user must purchase serial printer to
print titration result, titration data, titration curve etc. It is allowed to select ‘Model TP-16,
TP-24 or TP-40 serial printers (to print respectively 16 alphabets, 24 alphabets, 40 alphabets
each line). Since the width of printers is different, the printing form of the printers is obviously
different. We suggest Model TP-40 serial printer be available.


Note : The Baud rate of serial printers are 9600bps, N, 8, 1. The way of shake hands is K4=ON












Diag.56 Set Printer Type

View Titrate Mode Set Calibrate

Set current printer









Tp-16s Printer
Tp-24s Printer

Titrate

Tp-40s Printer

Clean

Enter

Exit

Feed

Titration Panel READY + K 10/21


User could just select the printer connected with the meter. When selection finished, press ‘Enter’ to
return to the starting status.

10.4 Set Burette

Two kinds of burettes are provided with the meter, 10ml and 20ml burette. The burette must be
set corresponding to the operating burette, otherwise, the meter will not make titration
correctly.
In starting status, press ‘Set’ Button in menu, select ‘Burette’ and press ‘Enter’ to set burette
type and coefficient. Each burette is marked with burette coefficient. User must set correctly
otherwise it will influence the titration result directly.











Diag. 57 Set burette and burette coefficient

View Titrate Mode Set Calibrate

Set Burette

Burette : 10ml burette

Titrate

Burette Coefficient: 100.00%

Clean

Enter

Exit

Feed


Titration Panel READY + K 10/21

42