LABEC M-TIT-5 User Manual

Page 49

Advertising
background image

48

Pre- titration has found false
end point

Setup of pre- titration
parameter is not suitable

Set jump to "large" according
to 7.1.5

Mode titration is not right
a. find false end point
b. can't find end point

a. pre- titration has found false

end point

b. mode is selected by mistake

a. turn off the false end point

according to 7.3.7

b. select the correct titration

mode

Preset end point titration is not
right
a. when over two end points,

after parameter is set,
titration can't be made

b. when titrating, display

"pre-controlled point setup
is not right"

a. parameter is set by mistake
b. parameter is set by mistake

a. set the correct parameter

according to 7.2.3

b. set the correct pre-

controlled point according
to 7.2.3.

Stirrer doesn't work

a. stirrer is not connected
b. stirring is set by mistake
c. stirrer is damaged
d. stirring drop has not been

put in the beaker

a. connect the stirrer according

to 1.2e in chapter IV
Operation of the Meter

b. speed up the stirrer

according to 10.1

c. change the stirrer
d. put stirring drop in the

beaker

There is bubble in feeding tube Leak solution on the joint of

feeding pipe

Install feeding tube according
to 1.1 c in chapter IV
Operation of the Meter

Mechanical part does not work
normally

Burette is not installed
correctly

Install burette according to
1.1b in chapter IV Operation
of Meter

Electrode is calibrated by
mistake

a. pH electrode's performance

is not right

b. buffer solution is made by

mistake

a. change pH electrode
b. make buffer solution again