7.4.4 End Point Jump Value Parameter

The same as End Point Jump Value Parameter in Pre-Titration Mode.


7.4.5 Pre-adding Volume Parameter

The same as Pre-adding Volume Parameter in Pre-Titration Mode.


7.4.6 Each Adding Volume Parameter

Blank titration mode is to add pre-adding volume first. When first adding finished, add the
same volume each time in sequence. The adding volume every time is the adding volume
after pre-adding volume.


7.4.7 End Volume Parameter

The same as Titration Mode End Volume Parameter
When all above setting finished, press ‘Start Titration’ button to start blank titration.


7.5 Manual Titration Mode

It can be used to add titrant manually and to make titration manually. This titration mode will
help user to find titration end point. In starting status, press ‘Titration’ button, select ‘Manual
Titration’ to enter into manual titration mode. Displayed as follows:

Manual Titration/Parameter setting










Start Titration


MV Titration pH Titration

Pre-adding: 10mL

Next Adding: 0.2mL


First end point

Jump Value: Large

Middle

Small

Value: 100

End volume: 40mL

MV/pH Titration

Jump Value Set

Stirring

Resume Set














Diag.36 Parameter Setting in Manual Titration Mode

All the same, Manual Titration Mode has following parameter: mV/pH Titration Parameter,

Pre-adding Volume Parameter , Next Adding Volume Parameter ,End Point Jump Parameter,

End Volume Parameter.

7.5.1 mV Titration Parameter

The same as mV/pH Titration Parameter in Pre-Titration Mode


7.5.2 Pre-adding Volume Parameter

The same as Pre-adding Volume Parameter in Blank Titration Mode.