Ambient conditions, Notice – VACUUBRAND MZ 2C NT +AK+M+D User Manual

Page 14

Advertising
background image

page 14 of 74

fore setting the voltage selection switch.

Note: If the

pump is switched on with wrong voltage selection, the

motor may be damaged!

• Ensure that the

coolant outlet pipe is always free and

that it cannot get blocked. If installing an optional cool-

ant valve, it must always be in the supply line of the

exhaust waste vapor condenser.

Make sure ventilation is adequate to maintain recommend-

ed operating temperature. Keep a minimum distance of

2 in (5 cm) between the cooling fan and surrounding items

(e.g., housing, walls, etc.), or else install an external auto-

matic ventilation system. Check fan regularly for dust/dirt.

Clean fan guard grill if necessary to avoid a reduction of

ventilation.

Use only hoses at the inlet and outlet of the pump with

an inner diameter at least as large as the diameter of the

pump’s tubing (to avoid overpressure at the outlet, and

reduction of pumping speed at the inlet).

Allow the equipment to equilibrate to ambient temperature

if you bring it from cold environment into a room prior to

operation. Notice if there is water condensation on cold

surfaces.

Secure coolant hoses at the hose nozzles (e.g., with hose

clamp) to prevent their accidental slipping.

Comply with all

applicable and relevant safety require-

ments (regulations and guidelines). Implement the re-

quired actions and adopt suitable safety measures.

Ambient conditions

Do not reach for this product if it has fallen into liquid.

There is a risk of deadly electrical shock. Unplug the

system immediately.