When assembling, ensure

vacuum-tightness. After as-

sembly, check the whole system for leaks.

Secure hose connections at the pump appropriately, e.g.,

with hose clamps, to protect against accidental detach-

ment.

To reduce pump noise emanating from the pump exhaust

port, connect an exhaust hose or use a silencer (see ”Ac-

cessories”, pg. 48).

Separator (inlet) and

exhaust waste vapor condenser (outlet)

Assembling the hose nozzle with union nut:

Take the hose nozzle with attached compression fer-

rule and union nut out of the catchpot and put onto in-

let connection (version 2 AK: onto inlet and outlet con-

nections).

Tighten the union nut by hand until you can feel the

stop. Then tighten an additional 1/4 rotation with an

open-ended wrench (size 17mm) for final installation.

Catchpots:

The catchpot at the inlet protects against drop-

lets and particles from entering the pump.
+

Enhances lifetimes of diaphragms and

valves.

+

Improves vacuum performance in applica-

tions with condensable vapors.

Both catchpots are coated with a protective

layer to protect against shattering in case of

breakage or implosion.

Assemble the catchpots at the inlet and at

the outlet using joint clips.

Exhaust waste vapor condenser:

Assemble the hose nozzles for coolant inlet

(12) and coolant outlet (13) tubing at the ex-

haust waste vapor condenser.

The

exhaust waste vapor condenser en-

ables an efficient condensation of the pumped

overpressure

safety relief