Use and operation, Installing a pump in a vacuum system – VACUUBRAND MZ 2C NT +AK+M+D User Manual

Page 38

Advertising
background image

page 38 of 74

Installing a pump in a vacuum system

If dangerous or polluting fluids could be released at

the outlet, install an appropriate system to catch and

dispose of those fluids.

+

Connect a gas-tight exhaust line at the pump outlet if

necessary. Always vent exhaust gases appropriately

(e.g., into a fume hood).

+

Never block or restrict the gas outlet. The exhaust line

must always be free of obstructions (no back pressure)

to ensure an unimpeded discharge of gas. The cross-

section of the outlet tubing must be at least the size of

the pump’s exhaust connection.

+

Particles and dust must not enter the pump. If nec-

essary, you must install appropriate filters. You must

ensure their suitability concerning gas flow, chemical

resistance and resistance to clogging prior to use.

+

Make sure ventilation is adequate, especially if the pump

is installed in an enclosure, or if the ambient temperature

is elevated. Provide external ventilation, if necessary.

• Reduce the transmission of vibration. Prevent mechan-

ical load due to rigid pipelines. Insert elastic hoses or

flexible elements as couplings between the pump and

rigid pipes.

Note: Flexible elements will compress or flatten when

evacuated if not designed for use under vacuum.

• Hose connections at the pump inlet must always be

gas tight.

• A power failure may cause accidental ventilation of the

pump, especially if the gas ballast valve is open. If this

constitutes a potential source of danger, take appropri-

ate safety measures.

• Check the power source and the pump’s rating plate

Use and operation

Advertising