Notice – VACUUBRAND MZ 2C NT +AK+M+D User Manual

Page 43

Advertising
background image

page 43 of 74

etc.), let the pump run with

gas ballast to help purge any

condensation in the pump.

Prevent internal condensation, transfer of liquids or dust.

The diaphragms and valves will be damaged, if liquids are

pumped in significant amounts.

Check the pump regularly for external soiling and depos-

its. Clean the pump if necessary to avoid an increase of

the pump’s operating temperature.

Operation with silencer (optional) at the outlet: Operating

the pump at a high inlet pressure or pumping dusty gases

for a long time may cause clogging of the silencer. Check

the silencer regularly and replace if necessary.

In case of overload, the motor is shut down by a

self-hold

thermal circuit breaker in the winding.

Note: Only a manual reset is possible. Switch off the

pump and disconnect the electrical power cord. Identify

and eliminate the cause of failure. Wait approximately five

minutes before restarting the pump.

Note: In case of supply voltage below 100V, the lock

of the breaker may not latch and the pump might re-

start on its own after sufficient cooling. Take appropri-

ate precautions, if an automatic restart of the pump

may lead to a dangerous situation.

A warm up period (approximately 15 min.) is required to

ensure that the rated ultimate vacuum and pumping speed

are attained. Avoid overheating (e.g., due to hot process

gases).

Pumps with flow control diaphragm valve:

Use the flow control diaphragm valve (14) at the pump

inlet to control the pumping speed. Open flow control dia-

phragm to pump down.

Opening the flow control diaphragm: Turn counter-

clockwise. Do not attempt to open the valve further

than when resistance is first encountered.

NOTICE