Important notes concerning the operation of the, Exhaust waste vapor condenser – VACUUBRAND MZ 2C NT +AK+M+D User Manual

Page 45

Advertising
background image

page 45 of 74

- Use inert gas for gas ballast to avoid the formation of

explosive mixtures. A special adapter fitting is needed

to connect an inert gas supply line (see ”Accessories”,

pg. 48). This adapter replaces the standard gas bal-

last cap and allows for an inert gas line to be connect-

ed via a KF DN 16 small flange at a maximum supply

pressure of 17.5 psi (1.2 bar) absolute.

- Close the gas ballast valve by turning the cap 180°.

In case of low boiling solvents (when the formation of

condensate is unlikely), the use of gas ballast might be

unnecessary. Operating the pump without gas ballast in-

creases the solvent recovery rate at the exhaust waste

vapor condenser.

Important notes concerning the operation of

the exhaust waste vapor condenser

Connect the exhaust to a suitable treatment system to

prevent the discharge of dangerous gases and vapors

to the surrounding atmosphere.

+

Never block the gas outlet ((2) hose nozzle for tubing

I.D. 3/8” (10 mm)). The exhaust hose must always be

unobstructed and without back-pressure to enable an

unhindered discharge of gases.

+

Check the overpressure safety relief device (11) at the

exhaust waste vapor condenser (10) regularly; replace

if necessary. Check especially for deterioration, co-

alescence and cracks.

• Ensure that the

coolant outlet hose is always free

and that it cannot get blocked.

• Maximum permissible coolant pressure at the exhaust

waste vapor condenser: 87 psi (6 bar) absolute

• Comply with the maximum permissible coolant pres-

sures of additional components in the coolant circuit

(e.g., coolant valve).