In normal use, the lifetime of the diaphragms and valves

is typically 15,000 operating hours. Bearings have a typi-

cal durability of 40000 h. Motor capacitors have a typi-

cal durability in the range of 10000 to 40000 h depending

strongly on operation conditions including ambient tem-

perature, humidity or load.

Ensure that the pump cannot be operated acciden-

tally. Never operate the pump if covers or other

parts of the pump are disassembled.

Switch off the pump. Disconnect the electrical

power cord and wait five seconds before starting

maintenance to allow the capacitors to discharge.

Note: The pump may be contaminated with process

chemicals, which have been pumped during operation.

Ensure that the pump is completely decontaminated

before maintenance commences.

+

Take adequate precautions to protect people from the

effects of dangerous substances if contamination has

occurred. Use appropriate protective clothing, safety

goggles and protective gloves.

+

Wear parts have to be replaced regularly.

+

Never operate a defective or damaged pump.

+

Check every motor capacitor regularly by measuring

its capacity and estimating its time in operation. Re-

place old capacitors early enough to prevent a failure

in operation. If an old motor capacitor fails, the capaci-

tor may get hot. It may even melt or emit a flame, which

could be

dangerous for persons and equipment in

the vicinity. The capacitors have to be replaced by an

electrician.

+

Vent the pump before starting maintenance. Isolate

the pump and other components from the vacuum sys-

tem. Allow sufficient cooling of the pump. Separate the