• We strongly recommend installing an optional coolant

valve (see ”Accessories”, pg. 48)

in the supply line

of the exhaust vapor condenser to save water and re-

duce the risk of water spill.

• Avoid overpressure in the coolant circuit (e.g., caused

by blocked or kinked coolant hoses).

In case of

condensation: Check the liquid level in both

catchpots (9) during operation. Check the liquid level in

both catchpots regularly. Do not allow the catchpots to

overfill. Drain catchpots in time to avoid overflow. Install a

level sensor (see ”Accessories”, pg. 48) for monitoring,

if necessary (VACUUBRAND controller CVC 3000 or VNC

2 is required).

The maximum liquid level is at approximately 80% of the

total filling level to avoid problems when removing the

catchpots.

Permissible range of coolant temperature at the exhaust

waste vapor condenser: 5 °F to 68 °F (-15°C to +20°C)

Check hose connections prior to starting operation of the

cooling system.

Check coolant hoses regularly during operation.

Removing the catchpots:

Catchpot at outlet:

Remove joint clip. Remove catchpot and drain conden-

sate.

Catchpot at inlet:

Admit air or inert gas (via the pump inlet) to restore at-

mospheric pressure in the catchpot before attempting

removal. Remove joint clip. Remove catchpot and drain

condensate.

Reattach drained catchpots.