Make sure ventilation is adequate to maintain recommend-

ed operating temperature. Keep a minimum distance of

2 in (5 cm) between the cooling fan and surrounding items

(e.g., housing, walls, etc.), or else install an external auto-

matic ventilation system.

Use connecting hoses with large diameter and keep them

as short as possible to avoid flow losses. Locate the pump

as closely as possible to the application.

Always install outlet tubing descending from the pump or

provide other measures to avoid backflow of condensate

towards the pump.

Use a suitable valve to isolate the pump from the vacuum

application. This is to allow the pump to warm up before

pumping condensable vapors and to clean the pump after

use before it is switched off.

NOTICE

Voltage selection switch:

1. Disconnect the electrical power cord.

2. Use a screw driver to adjust the

voltage selection

switch at the terminal box of the pump to the supply

voltage:

”115/120” corresponds to 90-126 V and

”230/240” corresponds to 180-253 V.

voltage selection

switch

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

Pump with dual-voltage motor: Check that the volt-

age selection switch at the terminal box is positioned

correctly. Check every time before starting the pump.

Note: If the pump is switched on with wrong voltage

selection, the motor may be damaged!

Change the selection at the voltage selection

switch only, if the pump is unplugged from the

power source.