To chose a pressure value for automatic switching off is
only possible if automatic switching off is activated (in mode
vacuum control with automatic).
➨ Press key MODE simultaneous with ▼ (arrow down).

☞ When the key is pressed shortly, the current pressure for

automatic switching off is displayed for one second (”AE”, i.
e. automatic switching off is preselected at the factory). With
a second tip within one second or continuous pressing:

➨ Select pressure for automatic switching off by using keys ▲

or

▼ (pressure selectable in a range from 1 to 1060 mbar or

”AE”; ”AE”, i. e. automatic switching off is active if a pres-
sure value < 1mbar is selected).

☞ The pressure is stored as new pressure for automatic switch-

ing off when the key is released.

Automatic switching off

MODE

30

mbar

Changing to mode continuous pumping

➨ Press key MODE simultaneous with ▲ (arrow up).

☞ Switching to continuous pumping is only possible, if vacuum

control is not activated (after pressing key START/STOP).

MODE

Advertising