➨ Check for washers. Do not mix the washers from the differ-

ent heads. Make sure that the original number is reassem-
bled at the individual pump head.

☞ Smaller number of washers: The pump will not attain final

vacuum. More washers: Clamping disc will hit head cover;
noise or even blockade of the pump.

➨ Position new diaphragm between diaphragm clamping disc

with square head screw and diaphragm support disc.

Note: Position diaphragm with white PTFE side to diaphragm

clamping disc (to pump chamber).

Replacing the diaphragm

➨ Use face wrench to remove diaphragm support disc.

➨ Remove head cover from housing cover insert and check

valves. Note position of valves and remove.

☞ Replace valves if necessary.

➨ Use petroleum ether or industrial solvent to remove depos-

its. Do not inhale.

➨ Check diaphragm for damage and replace if necessary. Use

Phillips screw driver to remove four countersunk head screws
and lift off housing plate.

Use the face wrench with torque indicator to assemble

diaphragm clamping disc, diaphragm and diaphragm sup-
port disc (and eventually washers) to the connecting rod.

☞ Make sure that the square head screw of the diaphragm

clamping disc is correctly seated in the guide hole of the
diaphragm support disc.

☞ Optimum torque for the diaphragm support disc: 6 Nm.

☞ The optimum torque is achieved if the pointer in the handle

of the VACUUBRAND face wrench shows to the longer mark-
ing line.

Assembling pump heads

➨ By turning eccentric bushing (front of connecting rod), bring

connecting rod into a position in which diaphragm is in con-
tact with housing and centred with respect to bore.