VACUUBRAND MZ 2C VARIO User Manual

Page 43

Advertising
background image

page 43 of 51

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999034 / 22/04/2008

➨ Use hex key to remove four socket head screws from pump

head and remove upper housing (housing cover with hous-
ing cover insert and head cover).

☞ Never remove parts by using a spiky or sharp-edged tool

(e.g. screw driver), we recommend to use a rubber mallet or
compressed air (to be blown carefully into port).

➨ Use open ended wrench (w/f 14) to turn elbow fitting 1/4 of a

turn, remove hose. Do not remove the elbow fitting from the
pump head.

☞ Through reassembly a leak may result.

View of the disassembled pump head parts

Diaphragm support disc

Diaphragm clamping disc
with connecting screw

Housing cover with
housing cover insert

Housing

Connecting rod

Washer

Diaphragm

Head cover

O-rings

Valves