Montage of the connecting hose

Individual performance check of a pump head

By measuring the pressure at the inlet port of the individual
head:
Use a suitable vacuum gauge (e. g. DVR 2, cat. no.: 682902),
make sure that it is correctly calibrated, and measure the pres-
sure at the inlet port. A vacuum of less than 120 mbar should
be indicated.
☞ If the reading is higher, recheck the pump chamber and make

sure that the valves and the diaphragms are correctly seated
(diaphragms concentric with bore).

➨ Use open ended wrench (w/f 14) to reconnect hose to elbow

fitting.

➨ Tighten union nuts first by hand and then tighten one full

turn using open ended wrench (w/f 17).

➨ Screw in four socket head screws fixing housing cover cross-

wise (e. g. in the sequence

➀, ➁, ➂, ➃) first slightly, then

tighten.

☞ Do not tighten until head cover is in contact with housing,

torque 12 Nm.

Reassemble in reverse order.
➨ Install head cover (with O-ring if applicable), valves and hous-

ing cover.

☞ Make sure that the valves are correctly seated: Valves at

the outlet with round centred opening under valve, valves at
the inlet with kidney-shaped opening beside valve.

➨ By turning eccentric bushing, bring connecting rod into up-

per turning point position (Max. stroke of the rod).

☞ Pay attention that the diaphragm stays positioned centrally

so that it will become clamped uniformly between housing
and head cover.