VACUUBRAND MZ 2C VARIO User Manual

Page 6

Advertising
background image

page 6 of 51

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. manual-no.: 999034 / 22/04/2008

Due to the residual leak rate of the equipment, there may be an exchange of gas,
albeit extremely slight, between the environment and the vacuum system.
☞ Adopt suitable measures to prevent contamination of the pumped substances or

the environment.

Pumping at high inlet pressure may lead to overpressure at the gas ballast valve.
☞ Pumped gases or condensate might be pushed out in case the valve is open.

☞ If an inert gas supply is connected, ensure that the inlet pipeline is not contami-

nated.

The controller is equipped with a short circuit proof transformer with an integrated
overload protection (no fuses).

☞ Failure of the pump (e. g. by power failure) or connected components, parts of the

supply (e. g. coolant) or change of parameters (e. g. increase of pressure in the
coolant system) must not lead to a critical dangerous situation under any circum-
stances.

Electronic equipment is never 100% fail-safe. This may lead to an indefinite status of
the equipment. Provide protective measures against disfunction and failure.
☞ Operating the pump with high or low frequency, stand still of the pump or operating

the venting valve must not lead to a critical dangerous situation under any circum-
stances.

Ensure that in case of failure the pump and the vacuum system always will turn into a
safe status.
☞ In case of diaphragm cracks or leaks in the manifold pumped substances might be

released into the environment or into the pump housing. To reduce the risk of leaks,
ask for a diaphragm pump with additional safety diaphragm.

☞ Comply especially with notes on operation and use and maintenance.

Use only genuine spare parts and accessories.
☞ Otherwise safety and performance of the equipment as well as the electromagnetic

compatibility of the equipment might be reduced.
Possibly the CE mark becomes void if not using genuine spare parts.

Ensure that maintenance is done only by suitable trained and supervised techni-
cians.
Ensure that the maintenance technician is familiar with the safety procedures
which relate to the product processed by the vacuum system and that the equipment,
if necessary, is appropriately decontaminated before starting maintenance.
Comply with local and national safety regulations.

Wear parts have to be replaced regularly. In case of normal wear the lifetime of the
diaphragms and valves is > 10000 operating hours. Bearings have a typical durability
of 40000 h.

Isolate equipment from mains and wait two minutes before starting mainte-

nance to allow the capacitors to discharge.

➨ Before starting maintenance, wait two minutes after isolating the equipment from

mains to allow the capacitors to discharge.

Ensure that the pump cannot be operated accidentally. Never operate the pump

if covers or other parts of the pump are disassembled. Never operate a defec-
tive or damaged pump.

Attention: In case of supply voltage below 100V, the lock of the cutout might be

restricted and the pump might restart on its own after sufficient cooling down. Take
suitable precautions, if an automatic restart of the pump may lead to a critical
dangerous situation.