VACUUBRAND PC 3012 VARIO plastic housing User Manual

Page 104

Advertising
background image

page 104 of 129

Detach the coupling of the connection tube (V)

to the other side of the pump at the valve head

(O).

PC 3002/3003/3004:

Detach the coupling of the connection tube (V)

to the other side of the pump as well as the

hose connection to the inlet/outlet of the vacu-

um system at the valve head (O).

V

O

U

Remove the 4 screws affixing the head cov-

er cowling (W) with a Torx driver T20. Pay

attention to the washers under the screws

and remove.

Pull off head cover cowling (W) carefully. Do

not tilt.

W

Disassemble the exhaust vapor condenser

EK:

Loosen the union nut at the inlet of the ex-

haust vapor condenser.

Remove the 4 screws affixing the counter

holder of the exhaust vapor condenser with

a Torx driver T10. Remove the exhaust va-

por condenser. In doing so pull the PTFE

hose out of the inlet of the exhaust vapor

condenser.

For maintenance, lay the pump / pumping unit on its side with the

pump heads to be maintained at the top. Support the pump / pumping

unit appropriately.

PC 3002/3003/3004 VARIO

Remove catchpots (14) at inlet and outlet

(see ”Use and operation”, pg. 38).

Advertising