VACUUBRAND PC 3012 VARIO plastic housing User Manual

Page 20

Advertising
background image

page 20 of 129

+

Make sure that the exhaust pipeline cannot become

blocked.

+

Check the overpressure safety relief device at the ex-

haust waste vapor condenser at appropriate intervals.

+

Attention: At pressures above approximately 795 Torr

(1060 mbar) the pressure reading becomes incorrect

due to saturation of the pressure transducer. The dis-

play flashes. Release pressure immediately! Risk of

bursting!

+

Comply with applicable regulations when disposing of

chemicals. Take into consideration that chemicals may

be contaminated. Take adequate precautions to pro-

tect people from the effects of dangerous substances

(chemicals, thermal decomposition products of fluoro-

elastomers). Use appropriate protective clothing and

safety goggles.

+

Interruption of the pump (e.g., due to power failure),

failure of connected components or of parts of the sup-

ply, or change in parameters must not be allowed to

lead to dangerous conditions. In case of a diaphragm

failure or in case of a leak in the manifold, pumped

substances might be released into the environment or

into the pump housing or motor.

Comply with all notes regarding proper use of the

pumps, as well as operation and maintenance guid-

ance.

+

The residual

leak rate of the equipment might render

possible an exchange of gas, albeit extremely slight,

between the environment and the vacuum system.

Adopt suitable measures to prevent contamination of

the pumped substances or the environment.

• Ensure that no parts of your clothing, hair or fingers

can be caught or drawn in at the inlet of the pump.

Never insert fingers or drop any other object into the

inlet or outlet.

Advertising