Shutdown & storage, Notice – VACUUBRAND PC 3012 VARIO plastic housing User Manual

Page 48

Advertising
background image

page 48 of 129

Check hose connections prior to starting operation of the

cooling system.

Check coolant hoses regularly during operation.

Removing the catchpots:

Catchpot or collecting flask at outlet:

Remove joint clip. Remove catchpot

or collecting flask and

drain condensate.

Catchpot at inlet:

Admit air or inert gas (via the pump inlet) to restore at-

mospheric pressure in the catchpot before attempting

removal. Remove joint clip. Remove catchpot and drain

condensate.

Catchpot at inlet (PC 301x VARIO):

Admit air or inert gas (via the pump inlet) to restore atmo-

spheric pressure in the catchpot before attempting remov-

al. Open drain screw (19) and drain condensate. Close

drain screw.

Reattach drained catchpots.

+

Important: Comply with regulations when disposing

of solvents/condensates. Recycle if possible; purify if

contaminated.

Shutdown & storage

The pump can be switched off under vacuum.

Short-term:

Has the pump been exposed to condensate?

- Allow the pump to continue to run at atmospheric pres-

sure for a few minutes.

NOTICE

NOTICE

Advertising