Vacuum controller cvc 3000 – VACUUBRAND PC 3012 VARIO plastic housing User Manual

Page 50

Advertising
background image

page 50 of 129

Vacuum controller CVC 3000

When switching on the controller CVC 3000 for the very first time, a menu

to select the language of the controller menu is displayed. Select the de-

sired language, e.g., ”English” by turning the selection knob and pressing

to confirm. Then select the pressure unit (”mbar”, ”Torr” or” hPa”) in the

same way.

It is possible to access the language selection menu at any time

by switching on the controller while keeping the selection knob

pressed.

After switching on the device, the

version number of the software is

displayed, followed by the preselected function and the pressure reading.

Attention: Do not assemble or remove plug connections off-axis!

Orient the plug correctly before inserting. To connect additional

components use VACUU•BUS Y-adapters and extension cables. If

an external pressure transducer is connected, it is recognized auto-

matically. Further information on how to use several sensors simul-

taneously is available upon request.

Keys

Start or Stop the process

VENT:

• A short tap vents momentarily;

process continues.

• Pressing longer than 2 seconds vents the

system to atmospheric pressure (to 788 Torr

(1050 mbar) at maximum); process stops.

MODE:

• Selects menu ”function”

• Use for temporary switching during

operation to other functions

ON/OFF switch

Advertising