VACUUBRAND PC 3012 VARIO plastic housing User Manual

Page 117

Advertising
background image

page 117 of 129

Loosely screw in the Allen head screws (R)

of the housing cover in a star pattern with a

5 mm wide Allen key, then tighten.

+

Recommended torque: 8.9 ft

.

lb

f

(12 Nm).

Check the performance by measuring the vacuum at the inlet port of the

individual pump head:

Use a suitable vacuum gauge (e.g., DVR 2, cat. no.: 682902), being sure

that it is correctly calibrated. Measure the vacuum at the inlet port (marked

IN). A vacuum of less than 120 mbar should be achieved.
+

If the reading is higher, recheck the pump chamber and make sure

that the valves and the diaphragms are correctly seated (diaphragms

concentric with bore).

Individual performance check of a pump head:

Replace diaphragms and valves of all eight pump heads

in the same way.

R

Replacing the valve at the distributor (outlet side)

Unscrew the two countersunk head screws

(M) at the manifold cover with an Phillips

screw driver. Remove cover plate and man-

ifold cover.

Loosen the union nut of the hose running directly to the manifold cover

(L) of the distributor (P) with an open-ended wrench (w/f) 17. Loosen

only the union nut at the pump head, not the one at the distributor.

Turn the fitting (Q) with an open-ended wrench (w/f 14) to detach the

hose from the pump head (1/4 of a turn at maximum).

+

Do not remove the elbow fitting from the pump head; during reassem-

bly a leak may result.

M

Advertising