VACUUBRAND PC 3012 VARIO plastic housing User Manual

Page 42

Advertising
background image

page 42 of 129

Catchpots:

The catchpot at the inlet protects against droplets and par-

ticles from entering the pump.

+

Enhances lifetimes of diaphragms and valves.

+

Improves vacuum performance in applications with

condensable vapors.

PC 300x VARIO:

Both catchpots are coated with a protective

layer to protect against shattering in case of

breakage or implosion.

Assemble the catchpots at the inlet and at

the outlet using joint clips.

PC 301x VARIO:

Assemble the inlet catchpot (18) with cen-

tring ring and clamping ring at the inlet of

the pump.

Assemble the collecting flask (20) at the

exhaust waste vapor condenser using the

joint clip.

In addition to connecting the pump inlet to

the application, install a vacuum connection

(hose) between the pressure transducer

(measurement connection at the rear of the

CVC 3000) and the vacuum system.

Exhaust waste vapor condenser:

Assemble the hose nozzles for coolant inlet

(15) and coolant outlet (16) tubing and for

the gas outlet (8) at the exhaust waste va-

por condenser.

The

exhaust waste vapor condenser en-

ables an efficient condensation of the pumped

vapors at the outlet.
+

No backflow of condensates.

+

Controlled recovery of condensates.

+

Close to 100% solvent recovery.

+

The isolation cover protects against glass

catchpot at

outlet (14)

catchpot

at inlet (14)

(18)

(19)

(15)

(16)

PC 300x VARIO

outlet (gas!) (8)

(15)

(16)

outlet (gas!) (8)

PC 301x VARIO

Advertising