Note the position and orientation of the valve

(B) and remove.

+

Check valve for damage and replace if nec-

essary. Make sure that the valve is correctly

seated.

Reassemble manifold cover (L) and cov-

er plate (K) and secure with countersunk

screws (M).

B

Notes on assembling fittings and hose connections

(only MV 10C VARIO-B, PC 3010 VARIO)

The thread of the fittings at the pump heads, at

the inlet and outlet distributors as well as the

hose connectors are sealed with PTFE tape.
+

If the fittings or the hoses have been re-

moved, use new PTFE bond (included in

set of seals) when reassembling.

Wind the thread of the fitting two times coun-

terclockwise with PTFE bond, the hose con-

nectors three times. Ensure that the PTFE

bond ends flush with the fittings.

hose connector

thread

Assembling the connecting hoses

Slip the hoses (O) onto the hose connec-

tors by turning the fittings (Q) with an open-

ended wrench w/f 14 (w/f 16 at outlet).

Tighten the union nuts (N) of the hose con-

nections at the pump heads with an open-

ended wrench w/f 17.

+

Tighten union nuts first by hand and then

tighten one full turn using the open ended

wrench.