VACUUBRAND MV 2 NT VARIO User Manual

Page 26

Advertising
background image

page 26 of 106

* Ultimate vacuum at setting ”Pump down” at speed ”HI”

** Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 1500rpm and ulti-

mate vacuum with standard silencer at outlet.

Type

MD 4 NT

VARIO

MV 2 NT

VARIO

MV 10

VARIO-B

Maximum pumping speed

50/60 Hz (ISO 21360)

cfm

(m

3

/h)

3.4

(5.7)

1.9

(3.3)

6.1

(10.4)

Ultimate vacuum

(absolute) *

Torr

(mbar)

0.75

(1.0)

0.2

(0.3)

0.2

(0.3)

Maximum permissible range of sup-

ply voltage ( ±10% )

Attention: Observe specifications of

rating plate!

100-120 V~ 50/60 Hz

-

200-230 V~ 50/60 Hz

Maximum rated current at:

100-120 V~ 50/60 Hz

200-230 V~ 50/60 Hz

A

A

6.3

2.5

-

3.5

Inlet

small flange DN 16

small flange

DN 25

Outlet

silencer

A-weighted emission

sound pressure level**

(uncertainty K

pA

: 3 dB(A))

dB(A)

45

45

52

Dimensions L x W x H

approx.

in

(mm)

12.8 x 9.3 x 9.6

(325 x 235 x 245)

19.1 x 8.7 x

11.9

(486 x 222 x

303)

Weight approx.

lbs. (kg) 41.0 (18.6)

41.7 (18.9)

55.1 (25.0)

Controller

CVC 3000

Pressure transducer

external gauge head VSK 3000,

ceramic diaphragm (alumina),

capacitive, absolute pressure,

gas type independent

Display

LCD graphic display, illuminated

Pressure units / scale (selectable)

Torr, mbar or hPa

Measuring range (absolute)

795 - 0.1 Torr (1060 - 0.1 mbar)

Maximum control range (absolute)*

795 - 0.1 Torr (1060 - 0.1 mbar)

Resolution (display)

0.1 Torr (mbar)

* The actual vacuum control range in your application might be reduced due to ultimate vacuum