Shutdown & storage, Notice – VACUUBRAND MV 2 NT VARIO User Manual

Page 37

Advertising
background image

page 37 of 106

In case of overload, the motor is shut down by a

self-hold

thermal circuit breaker in the winding.

Note: Only a manual reset is possible.

Press the START/

STOP key at the controller to confirm the error message.

Switch off the pump and disconnect the electrical power

cord. Identify and eliminate the cause of failure. Wait ap-

prox. five minutes before restarting the pump.

Check fan regularly for dust/dirt. Clean if necessary. Avoid

overheating (e.g., due to hot process gases).

A warm up period (approximately 15 min.) is required to

ensure that the rated ultimate vacuum and pumping speed

are attained.

Shutdown & storage

The pump can be switched off under vacuum.

Short-term:

Has the pump been exposed to condensate?

- Allow the pump to continue to run at atmospheric pres-

sure for a few minutes.

Has the pump been exposed to media which may damage

the pump materials or form

deposits?

- Check and clean pump heads if necessary.

Has the pressure transducer been exposed to media which

may form deposits?

- Clean pressure transducer if necessary.

Long-term:

- Take measures as described above regarding short-

term shutdown.

- Separate the pump from the application.

- Close inlet and outlet ports (e.g., with transport caps).

- Store the pump under dry conditions.

NOTICE